Difference between revisions of "Sphere lab technical info"

From biophysics
Jump to navigation Jump to search
 
(51 intermediate revisions by the same user not shown)
Line 7: Line 7:
 
*2018: all original Visaton speakers where replaced by Minx12 speakers
 
*2018: all original Visaton speakers where replaced by Minx12 speakers
 
*2018: Remmel head tracking system replaced by Femto
 
*2018: Remmel head tracking system replaced by Femto
 +
*2024: PA5's are not used anymore, scaling in WAV-file
  
 
==Sound booth==
 
==Sound booth==
Line 12: Line 13:
 
*Dimensions: LxWxH = 420x300x300cm
 
*Dimensions: LxWxH = 420x300x300cm
 
===Description===
 
===Description===
The sound booth is an acoustically isolated room with sound absorbing materials on all walls and the floor to reduce reverberation. A 2.75m diameter sphere build of metal tubes holds about 112 small passive speakers and 14 large active speakers. The small speakers also contain 5mm two color LEDs. In the centre of the sphere there is a chair for a subject. The chair is placed in a way that the head of the subject is right in centre of the sphere.  
+
The sound booth is an acoustically isolated room with sound absorbing materials on all walls and the floor to reduce reverberation. A 2.75m diameter sphere build of metal tubes holds about 112 small passive speakers and 14 large active speakers. The small speakers also contain 5mm two color LEDs. In the centre of the sphere there is a chair for a subject. The chair is placed in a way that the head of the subject is right in centre of the sphere. The positions of the speakers and LEDs are specified in [[Coordinate systems|Spherical coordinates]].
 
Large coils are embedded in the walls of the sound booth. The coils make a cube of about 3mx3mx3m. These coils are used for head movement detection.
 
Large coils are embedded in the walls of the sound booth. The coils make a cube of about 3mx3mx3m. These coils are used for head movement detection.
 
A metal cabinet contains PLC equipment for controlling the LEDs as well as a patch panel for the small speakers. The large speakers are connected with XLR cables to a large reel.
 
A metal cabinet contains PLC equipment for controlling the LEDs as well as a patch panel for the small speakers. The large speakers are connected with XLR cables to a large reel.
Line 22: Line 23:
 
*Floor: anti-fatigue rubber floor mat with holes
 
*Floor: anti-fatigue rubber floor mat with holes
 
see [[Acoustic Materials]]
 
see [[Acoustic Materials]]
 +
 +
===Coordinates===
 +
The orientation of the coordinates is determined by the position of the head of the subject looking to the center speaker. The raw acquisition data of the head coil results in tuples of voltages (Frontal, Vertical, Horizontal)
 +
 +
The FVH coordinate system relate to the medical RAS coordinate system (Right, Anterior, Superior) when looking to the center speaker in the following manner:
 +
 +
* Frontal <--> Anterior
 +
* Vertical <--> Superior
 +
* Horizontal <--> Right
 +
 +
For experiments the speaker positions are specified in Double Polar coordinates (Azimuth, Elevation, Hemisphere), where the hemisphere can be front or back.
 +
 +
For more information about the different coordinate systems see [[Coordinate systems]]
  
 
==Computer==
 
==Computer==
Line 28: Line 42:
 
==Software==
 
==Software==
 
The software consist of several parts:
 
The software consist of several parts:
*Matlab software for generating experiment files.
+
*GenExp: Matlab software for generating experiment files.
*Matlab software that runs the experiments.
+
*SpherePrime: Matlab software that runs the experiments.
 
*RPvdsX software for the RP2.1 soundprocessors
 
*RPvdsX software for the RP2.1 soundprocessors
 
*RPvdsX software for the RA16 medusa base station.
 
*RPvdsX software for the RA16 medusa base station.
Line 41: Line 55:
 
*genexp_student.m
 
*genexp_student.m
  
===Sphere Prime===
+
===NewSpherePrime===
The software mostly used is called 'SpherePrime'. It can be found in the '''\biofysica\experiment''' directory. It has a graphical user interface (GUI) and is, except for the GUI, written in the procedural programming paradigm. Nearly all functions have a parameter 'handles' as input parameter and as output parameter. 'handles' is a struct that contains nearly all information that is moved around in the program. Every function has the ability to change 'handles'. This way it acts as a global data structure.
+
The current program (since 2024) is 'newSpherePrime' developed by Ruurd Lof. It can be found in the '''\biofysica\experiment\newSpherePrime''' directory. This program has a GUI build in the app designer of Matlab. The program is mostly written in the Object Oriented Programming (OOP) paradigm. The core of the program is a state machine. Due to its structure the GUI stays always responsive.
  
A newer version of 'SpherePrime' is 'newSpherePrime'. This program also has a GUI build in the app designer of Matlab. The program is mostly written in the Object Oriented Programming (OOP) paradigm. The core of the program is a state machine. Due to its structure the GUI stays always responsive.
+
The 'newSpherePrime' program uses a vector based method for locating the stimuli, which is more precise than the 'interpolant' method.
 +
 
 +
===Older SpherePrime===
 +
The software mostly used up to 2023 is called 'SpherePrime'. It is written in the procedural programming paradigm. Nearly all functions have a parameter 'handles' as input parameter and as output parameter. 'handles' is a struct that contains nearly all information that is moved around in the program. Every function has the ability to change 'handles'. This way it acts as a global data structure.
 +
 
 +
The 'SpherePrime' program uses the 'interpolant' method for locating the stimuli. This can sometimes lead to different speaker choices for locations in the backward hemisphere.
 +
 
 +
<pre>
 +
N.B. SpherePrime uses the outdated GUIDE environment for the GUI. The GUIDE environment will be removed in a future releases of MATLAB.
 +
</pre>
  
 
===RPvdsEx software===
 
===RPvdsEx software===
Line 77: Line 100:
 
|
 
|
 
|-
 
|-
|Zbus
+
|Zbus *
|PA5 Attenuator
+
|PA5 Attenuator *
|PA5 Attenuator
+
|PA5 Attenuator *
|F05 optibit bus #1
+
|F05 optibit bus #1 **
 
|-
 
|-
|Zbus
+
|Zbus *
|PA5 Attenuator
+
|PA5 Attenuator *
|PA5 Attenuator
+
|PA5 Attenuator *
|F05 optibit bus #2
+
|F05 optibit bus #2 **
 
|-
 
|-
 
|Zbus
 
|Zbus
Line 122: Line 145:
 
|F05 optibit bus #4
 
|F05 optibit bus #4
 
|}
 
|}
 +
 +
* The PA5's are not used anymore. Loudness has to be scaled in the WAV files.
 +
** The optibit bus #1 and optibut bus #2 are disconnected.
  
 
The next list shows all other electronics in the lower half of the rack from top to bottom.
 
The next list shows all other electronics in the lower half of the rack from top to bottom.
Line 290: Line 316:
 
====Parts====
 
====Parts====
 
*Trigger from Medusa base station
 
*Trigger from Medusa base station
*PLC cabinet (houses PLC system)
+
*[[PLC LED controller specifications|PLC LED controller]]
**FP2 PLC system from Panasonic
+
*Break-out panel with 5 pole Binder connectors (128x)
**Break-out panel with 5 pole Binder connectors (128x)
 
 
*Binder cables (120x)
 
*Binder cables (120x)
 
*LED mounting frames (120x)
 
*LED mounting frames (120x)
*Red/Green LED’s (120x)
+
*[[LED specifications|Bivar 5BC-3-CA-F]] Red/Green LED’s (120x)
 
*Trigger converter (5V &rarr; 24V) for input trigger
 
*Trigger converter (5V &rarr; 24V) for input trigger
 
*Trigger converter (24V &rarr; 5V) for trigger echo
 
*Trigger converter (24V &rarr; 5V) for trigger echo
  
====Specifications Panasonic FP2 series====
+
 
See [[PLC LED controller specifications]]
 
  
 
====Specifications Red/Green LEDs====
 
====Specifications Red/Green LEDs====
See [[LED Specifications]]
+
See [[LED specifications]]
  
 
==Trigger/Timing system==
 
==Trigger/Timing system==
Line 313: Line 337:
  
 
===Parts===
 
===Parts===
*Pushbutton (with 9V battery for 5V output)
+
*[[Active handheld push button (5V output)|Active Pushbutton]] (with 9V battery for 5V output)
 
*Arduino [[Trigger distributor]] (5V inputs (3x), 5V outputs (6x), 3V outputs (2x), testbutton)
 
*Arduino [[Trigger distributor]] (5V inputs (3x), 5V outputs (6x), 3V outputs (2x), testbutton)
 
*PP16 patch panel for analogue inputs
 
*PP16 patch panel for analogue inputs
 
*PP16 patch panel for digital outputs
 
*PP16 patch panel for digital outputs
 
*Medusa Base Station (TDT RA16)  
 
*Medusa Base Station (TDT RA16)  
*[[Digital event recorder]]
+
*[[8 channel digital event recorder|Digital event recorder (LSLDER05)]]
  
 
===Specification===
 
===Specification===
Line 328: Line 352:
  
 
==Head Tracking System==
 
==Head Tracking System==
 +
[[File:SphereLab_Lockin_settings.JPEG|300px|thumb|right|Lock-in amplifier settings]]
 +
*See [[EM Field Head Tracking System specifications]]
  
The Head Tracking System is used for recording head movements of the subject during a trial. The system consists of the following parts:
+
===Lock-in amplifiers settings===
[[File:SphereHeadTrackingSystem.png|300px|thumb|right|Schematic of the Sphere Head Tracking System]]
 
*Three oscillators at roughly 50, 60 and 70 kHz for X, Y and Z direction.
 
*Three signal amplifiers that amplify the oscillator signals.
 
*six field coils that make a cube of roughly 3x3x3 meters.
 
*A small pick-up coil mounted on a glasses frame that is worn by the subject during the experiment.
 
*Three lock-in amplifiers that split the signal from the pick-up coil in the 3 frequency components and measures their envelope.
 
*An acquisition device that converts the output signals from the lock-in amplifiers.
 
*A RZ6 multi I/O processor that records the acquired data.
 
*Software that converts the recorded data to head orientations using calibration data.
 
  
{| class="wikitable"
 
! Channel
 
! Frequency
 
! Orientation
 
! Color
 
! Polarity
 
|-
 
| Channel 1
 
| 50 kHz
 
| Frontal
 
| Red
 
| Front = positive; Back = negative
 
|-
 
| Channel 2
 
| 60 kHz
 
| Vertical
 
| Blue
 
| Right = positive; Left = negative
 
|-
 
| Channel 3
 
| 70 kHz
 
| Horizontal
 
| Yellow
 
| Up = positive; Down = negative
 
|}
 
 
===Lock-in amplifiers===
 
[[File:SphereLab_Lockin_settings.JPEG|300px|thumb|right|Lock-in amplifier settings]]
 
Three Femto LIA-BV-150-H Lock-in amplifiers are used in order to separate the Frontal, Horizontal and Vertical signals from the pick-up coil. Each Lock-in amplifier has a Sinus Oscillator Module (SOM-1) that is used both as Field Modulation Signal and as Reference Signal. The Lock-in amplifier settings are set at the front of the device (see picture), except for the modulation/reference frequency which is set inside on the SOM-1-board.
 
  
 
{| class="wikitable"
 
{| class="wikitable"
 
! Setting
 
! Setting
! Frontal
+
! Frontal (Red)
! Vertical
+
! Vertical (Blue)
! Horizontal
+
! Horizontal (Yellow)
 
|-
 
|-
 
| Frequency
 
| Frequency
Line 412: Line 400:
 
|-
 
|-
 
| PLL
 
| PLL
| F
+
| S
| F
+
| S
| F
+
| S
 
|-
 
|-
 
| Reference Threshold
 
| Reference Threshold

Latest revision as of 13:17, 27 November 2024

back to Sphere lab

Introduction

The Sphere setup is a sound boot with about 126 speakers arranged in a sphere. In the middle of the sphere is a chair where a person (the subject) can sit, so that his/her head is exactly in the centre of the sphere. The subject can be presented with stimuli in the form of sounds or led flashes. The headmovements of the subject can be tracked. The subject can also respond to stimuli by pressing a button. The experiments are controlled by a computer and electronics from outside the boot.

History:

  • 2014: build of the original system
  • 2018: all original Visaton speakers where replaced by Minx12 speakers
  • 2018: Remmel head tracking system replaced by Femto
  • 2024: PA5's are not used anymore, scaling in WAV-file

Sound booth

Sound Booth
  • Dimensions: LxWxH = 420x300x300cm

Description

The sound booth is an acoustically isolated room with sound absorbing materials on all walls and the floor to reduce reverberation. A 2.75m diameter sphere build of metal tubes holds about 112 small passive speakers and 14 large active speakers. The small speakers also contain 5mm two color LEDs. In the centre of the sphere there is a chair for a subject. The chair is placed in a way that the head of the subject is right in centre of the sphere. The positions of the speakers and LEDs are specified in Spherical coordinates. Large coils are embedded in the walls of the sound booth. The coils make a cube of about 3mx3mx3m. These coils are used for head movement detection. A metal cabinet contains PLC equipment for controlling the LEDs as well as a patch panel for the small speakers. The large speakers are connected with XLR cables to a large reel. An infrared camera is installed in the booth. The experimenter has a monitor from which he/she can see the inside of the booth. The camera looks down on the back of the subject.

Acoustics

The walls and floor of the booth as well as the PLC cabinet and the leg rest are covered with sound dampening materials. The speaker frame sphere is of 1 cm thick steel tubes.

  • Walls: egg box type soundproofing foam
  • Floor: anti-fatigue rubber floor mat with holes

see Acoustic Materials

Coordinates

The orientation of the coordinates is determined by the position of the head of the subject looking to the center speaker. The raw acquisition data of the head coil results in tuples of voltages (Frontal, Vertical, Horizontal)

The FVH coordinate system relate to the medical RAS coordinate system (Right, Anterior, Superior) when looking to the center speaker in the following manner:

  • Frontal <--> Anterior
  • Vertical <--> Superior
  • Horizontal <--> Right

For experiments the speaker positions are specified in Double Polar coordinates (Azimuth, Elevation, Hemisphere), where the hemisphere can be front or back.

For more information about the different coordinate systems see Coordinate systems

Computer

zBus Monitor

A windows computer with MATLAB, RPvdsEx and zBUSmon. The computer has an optical interface card (PO5e) for communication with the Tucker Davis equipment via the Optibit optical bus (FO5) at the back of the rack. The working of the optical bus can be monitored by the program zBUSmon program from TDT. This program has also some control functions for the optical bus. It shows all the connected zBus chassis and the TDT devices that are installed in each chassis. It also shows the version number of the installed firmware on the devices.

Software

The software consist of several parts:

  • GenExp: Matlab software for generating experiment files.
  • SpherePrime: Matlab software that runs the experiments.
  • RPvdsX software for the RP2.1 soundprocessors
  • RPvdsX software for the RA16 medusa base station.

GenExp

The software that generates a experiment file is called 'genexp_xxxxx'. In the \biofysica\experiment\exp directory there are five examples:

  • genexp_defaultcal.m
  • genexp_defaultloc.m
  • genexp_fartloc.m
  • genexp_glausndloc.m
  • genexp_student.m

NewSpherePrime

The current program (since 2024) is 'newSpherePrime' developed by Ruurd Lof. It can be found in the \biofysica\experiment\newSpherePrime directory. This program has a GUI build in the app designer of Matlab. The program is mostly written in the Object Oriented Programming (OOP) paradigm. The core of the program is a state machine. Due to its structure the GUI stays always responsive.

The 'newSpherePrime' program uses a vector based method for locating the stimuli, which is more precise than the 'interpolant' method.

Older SpherePrime

The software mostly used up to 2023 is called 'SpherePrime'. It is written in the procedural programming paradigm. Nearly all functions have a parameter 'handles' as input parameter and as output parameter. 'handles' is a struct that contains nearly all information that is moved around in the program. Every function has the ability to change 'handles'. This way it acts as a global data structure.

The 'SpherePrime' program uses the 'interpolant' method for locating the stimuli. This can sometimes lead to different speaker choices for locations in the backward hemisphere.

N.B. SpherePrime uses the outdated GUIDE environment for the GUI. The GUIDE environment will be removed in a future releases of MATLAB.

RPvdsEx software

In the \biofysica\experiment directory there are several versions of .rcx programs. For the RP2.1 soundprocessors

  • sphere_RP2_GWN.rcx
  • sphere_RP2_WAV.rcx

For the RA16

  • sphere_RA16.rcx
  • sphere_RA16_equaltowaitevent.rcx
  • sphere_RA16_testled.rcx

When a RZ6 is used instead of a RA16 there is:

  • sphere_RZ6.rcx

Electronics rack

upper half of electronics rack

The upper half of the electronics rack contains 10 TDT Zbus devices chassis, each containing two TDT series 3 devices. The lower half contains the Field coil amplifier and the Femto LockinAmplifier as part of the head tracking electronics.

The following table shows all the zBus chassis from top to bottom, with the installed devices.

Chassis Front left Front right Back
Zbus MS2 Monitor speaker <empty>
Zbus * PA5 Attenuator * PA5 Attenuator * F05 optibit bus #1 **
Zbus * PA5 Attenuator * PA5 Attenuator * F05 optibit bus #2 **
Zbus SA1 Speaker amplifier SA1 Speaker amplifier
Zbus RP2.1 Sound processor RP2.1 Sound processor F05 optibit bus #3
Zbus PM2R Multiplexer PM2R Multiplexer
Zbus PM2R Multiplexer PM2R Multiplexer
Zbus PM2R Multiplexer PM2R Multiplexer
Zbus PM2R Multiplexer PM2R Multiplexer
Zbus RA8GA2 multi DAC RA16ba base station F05 optibit bus #4
* The PA5's are not used anymore. Loudness has to be scaled in the WAV files.
** The optibit bus #1 and optibut bus #2 are disconnected.

The next list shows all other electronics in the lower half of the rack from top to bottom.

  • PP16 Patch panel for Medusa Base Station (2x)
  • Digital event recorder
  • Junction box panel (connects to junction box in sound boot)
  • Trigger distribution panel
  • Femto Lockin amplifiers
  • Tektronics Oscilloscope
  • Field coil amplifier
  • Rack main power switch
  • Patch panel for XLR output to Tannoy speakers

Optical Interface

Optical interface card

Four of the Zbus chassis in the rack have an optical interface (F05) at the back. These optical interfaces are connected in a loop with the P05e interface at the back of the computer. PO5e optibit bus F05 optibit bus (4x) Transfer rate: read 1.5 Mbit/s, write 1.0 Mbit/s N.B. The optical bus is not 100% stable. Once in a while it hangs up the program. Issue #196 on GitLab.

Sound system

Connection schematics of the sound system
Speaker locations on the sphere

The sound system consists of two programmable DSP’s (TDT RP2.1’s) with each two DAC’s. An RP2.1 generates a sound signal that travels via a programmable attenuator (TDT PA5), a stereo amplifier (TDT SA1) and a multiplexer (TDT PM2R) to a patch panel inside the PLC cabinet in the boot and from there to a speaker. The multiplexing system consists of two sets of four multiplexers (TDT PM2relay). Each multiplexer has 16 channels. Only one channels per multiplexer can be opened at a time. The total number of channels is 128, but the numbering of the channels (and speaker IDs) is 0-127. Each RP2.1 controls four multiplexers. Since each RP2.1 has 2 DAC outputs, each output is connected to two multiplexers. It is therefore possible to play up to 4 different sounds at once over different speaker. Parts

  • TDT RP2.1 (2x)
  • TDT PA5 (4x)
  • TDT SA1 (2x)
  • TDT PM2relay Multiplexers (8x)
  • Flat cables from RP2.1 to PM2relays (2x)
  • Dsub25 cables from PM2relays to PLC cabinet in the boot (8x)
  • Patch panel inside the PLC cabinet
  • Speakers (116x)

Minx Min12 speaker system

Speaker (and LED) connection scheme

On the sphere frame Cambridge Audio Minx Min12 speakers are used.

Speaker specifications:

  • Sensitivity: 86 dB SPL (@2.83 Vrms input)
  • Frequency response: 150 Hz-20 kHz
  • Impedance: 8 Ohms
  • H x W x D: 78 x 78 x 85 mm
  • Weight: 0.43 kg

The speakers are connected via 5 lead wires with a 5 pole binder connector at each end. Two leads are used for the speakers, three leads are used for the LED’s.

  • RP2.1_1 Left : 26 speakers #000..#025
  • RP2.1_1 Right: 28 speakers #032..#057
  • RP2.1_2 Left : 32 speakers #064..#095
  • RP2.1_2 Right: 26 speakers #096..#123

A total of 112 speakers.

The channels numbers (or speakers IDs) that are not used are: [26 27 28 29 30 31 58 59 60 61 62 63 124 125 126 127]

Azimuths and elevations were measured by Sebastian Ausili in June 2015 and are tabulated in 'Sphere Measurement.xlsx'. The newSpherePrime program uses the same data in the file 'sphere_sound_locations.xlsx'.

Tannoy speaker system

XLR patch panel connection scheme

There are 14 Tannoy Reveal 402 speakers in the horizontal plane, positioned on small plateaus. These are active speakers with their own amplifier. The speakers are used when lower frequencies (50Hz-150Hz) are important. Due to hiss problems, the original block transformers have been replaced by toroidal transformers. The Tannoy speakers are connected with XLR cables to a special patch panel at the bottom of the electronics rack, and inside the boot to a panel on a XLR cable reel. Since the output of the TDT soundsystem does not have a balanced output, the shield of the XLR cables are connected to the minus leads. The Tannoy speakers can be connected to a multiplexer using a special flat cable that connects to the back of the XLR patch panel.

A flat cable with DSub25 connector from the back of the XLR patch panel you can connect the Tannoy speakers to a multiplexer. In the following table you see the mapping from the mux channels to the XLR channels. The collumn "Pin MUX" shows the corresponding DSub25 pin numbers.

Channel Pin MUX XLR
X 1 GND
A0 15 1
A1 3 2
A2 16 3
A3 4 4
A4 17 5
A5 5 6
A6 18 7
A7 6 8
A8 19 9
A9 7 10
A10 20 11
A11 8 12
A12 21 13
A13 9 14
X X 15
X X 16

The wiring of the XLR sockets on the patch panel is as follows:

  • XLR-1: Shield = zwart
  • XLR-2: Hot = rood
  • XLR-3: Cold = connected to XLR-1

Speaker specifications:

  • Active speaker: Bi-Amp (25W + 25W)
  • 4" Woofer; 3/4" Tweeter
  • Frequency range: 56 Hz tot 43 kHz
  • Cross over freq.: 2.8 kHz
  • Distortion: <0.9%
  • Maximum sound pressure 101 dB
  • Bass reflex system
  • H x W x D: 240 x 147 x 212 mm
  • Weight: 5.2 kg

LED system

PLC output scheme (only 2 outputs indicated)

The LED system consists of a cabinet inside the sound boot with a Panasonic FP2 series PLC system (programmable logic controller). The PLC can independently switch 256 channels. The channels connect to a two color LEDs (red and green) that are mounted on the speaker. The PLC cabinet has two test buttons that can light all LEDs at once. The PLC system consists of a PLC with several IO units and is can be programmed by via a special USB cable.

Parts

  • Trigger from Medusa base station
  • PLC LED controller
  • Break-out panel with 5 pole Binder connectors (128x)
  • Binder cables (120x)
  • LED mounting frames (120x)
  • Bivar 5BC-3-CA-F Red/Green LED’s (120x)
  • Trigger converter (5V → 24V) for input trigger
  • Trigger converter (24V → 5V) for trigger echo


Specifications Red/Green LEDs

See LED specifications

Trigger/Timing system

Scheme of trigger system
LED stimulus in standard modus (X=0)
LED stimulus in short pulse modus (X=1)

Description

The task of the trigger/timing system is to distribute trigger signals and record the timing of the triggers. In the standard situation when an experiment starts, the subject has to push a button to start a trial. The button push is registered by an arduino trigger bridge that outputs a trigger to the Medusa Base Station. The Medusa sends triggers to the RP2.1’s, the LED PLC and to the Event Recorder. The Event Recorder timestamps all the triggers and can be read out via ethernet. The arduino trigger bridge has a green LED that indicates that a trigger signal is generated and it has a test button that generates a trigger signal when pushed.

Parts

Specification

The LED PLC in the LED controller cabinet can be programmed by Matlab. The PLC stores a list of LED configurations. One configuration tells the PLC which LEDs should be on or off. In the standard mode (X = 0) the current configuration is changed to the next configuration by an electrical trigger signal from the Medusa Base Station. A stimulus of a single LED consists of two configurations, one with only that LED on, followed by one with all the LEDs off. The timing and the length of the stimulus are controlled by trigger signals sent from the Medusa Base Station. The PLC cycle of 1.35 ms is the limiting factor for the shortness of a LED stimulus. In practice the pulselength can be as short as two or three PLC cycles (2.7 ms or 4.0 ms) in the standard modus (X=0). The number of cycles the stimulus lasts can vary by one cycle. In practice there is always 1.35 ms jitter on top of a 1.35 ms delay for the start (and end) of a LED stimulus (see figure). When more precise knowledge is needed over the timing or length of the stimulus, a timing signal (trigger echo) sent from the PLC can be recorded. The trigger echo signal is lined up with the rising or falling flanks of the LED and can measure the real stimulus with an accuracy down to 40 µs.

When using short stimuli the X parameter can be used to control the exact number of cycles. When X = n the length of the LED stimulus is exactly n PLC cycles.

Head Tracking System

Lock-in amplifier settings

Lock-in amplifiers settings

Setting Frontal (Red) Vertical (Blue) Horizontal (Yellow)
Frequency 50 kHz 60 kHz 70 kHz
Time constant B B B
Sensitivity 4 5 5
Phase Course 0 F 4
Phase Fine 4 D 6
Dynamic Reserve L L L
Mode 1f 1f 1f
PLL S S S