Artinis Oxymon NIRS

Description
The Artinis Oxymon is a stationary, fibre-based functional near-infrared spectroscopy (fNIRS) system made by Artinis Medical Systems (Elst, the Netherlands). It measures changes in the concentration of oxygenated haemoglobin (O2Hb), deoxygenated haemoglobin (HHb) and total haemoglobin (tHb) in the outer layer of the cortex, using continuous-wave NIRS and the modified Beer-Lambert law.
The light sources and detectors are in the Oxymon cabinets. Optical fibres carry the near-infrared light to the head and back. The fibre ends (optodes) are placed in optode holders in the headcap. The system is modular: one cabinet gives at most 8 channels, and up to three cabinets can be combined into one system with 24 channels. Each cabinet can also be used as a smaller, independent system.
In the NIRS-EEG lab there are two 24-channel Oxymon systems. Each system has 8 transmitters and 4 receivers (split fibres). With two systems a large part of the head can be covered, e.g. both temporal lobes for auditory experiments.
Compared with the wearable Brite, the Oxymon has a higher sample rate (50 Hz as standard, up to 250 Hz) and analog inputs for triggers, but the subject is connected to the cabinets by the fibres and should sit still. The Oxymon is therefore used for experiments in the booth in which the subject sits in a chair.
Recording and synchronisation
The Oxymon is operated with OxySoft on the recording computer. OxySoft shows the raw optical densities and the haemoglobin concentrations in real time and stores the data. During subject preparation the signal quality is checked in OxySoft: a clear heartbeat (~1 Hz) should be visible in the raw signals of good channels. The optode templates for our layouts are in nirs/layout/optodetemplates.xml in the biofysica repository.
The Oxymon has 8 AD channels (analog inputs) to record triggers together with the NIRS data. According to Artinis, the AD channels are sampled at the same rate as the NIRS channels. If you record at a low sample rate, make the trigger pulses long enough that they are not missed. See NIRS-EEG technical info for how the triggers are routed in the lab.
Specifications
General values from the Artinis website, plus the configuration of the systems in the NIRS-EEG lab.
| Specification | Details |
|---|---|
| Technique | Continuous-wave NIRS (modified Beer-Lambert law) |
| Measured parameters | Changes in O2Hb, HHb and tHb; raw optical densities |
| Channels | up to 8 per cabinet; 4 to 24 per system |
| Configuration in the NIRS-EEG lab | 2 systems, each with 24 channels, 8 transmitters and 4 receivers (split fibres) |
| Wavelengths | 2 per channel (up to 4 possible) |
| Sample rate | 50 Hz standard, up to 250 Hz |
| Analog inputs | 8 AD channels (sampled at the NIRS sample rate) |
| Fibres | optical fibres, 3 m or 10 m, MRI-compatible versions available |
| Short-separation channels | possible |
| Software | OxySoft (Windows) |
| Synchronisation | AD channels; parallel sync cable / PortaSync |
| Compatible with | EEG, TCD, (f)MRI |