TDT Series 3
Description
The TDT System 3 was introduced in the late 1990. It is a versatile and widely used data acquisition and processing system in the field of neurophysiology and electrophysiology research. It's a hardware and software platform designed for acquiring, processing, and analyzing neural signals.
Key features of the TDT System 3 include:
- High-Speed Data Acquisition: It is capable of recording neural signals, such as multi-channel electrophysiological data, at high sampling rates.
- Flexible Signal Processing: The system provides a range of signal processing capabilities, including filtering, spike sorting, and event detection.
- Integration with MATLAB: MATLAB compatibility is a notable feature, making it easier for researchers like yourself who use MATLAB for data analysis and programming.
- Customizable: Researchers can customize and expand the system with additional modules and accessories to suit their specific experimental needs.
- Multi-Channel Capability: It can handle multiple channels simultaneously, which is crucial for studying neural activity in various brain regions.
System 3 devices
RP2.1 Enhanced Real-Time Processor
The RP2.1 is a single core Digital Sound Processor. It can be programmed by the RPvdsEx programming interface. It can generate signals or play signals from memory as well as record signals.
- Sample rates: 6, 12, 25, 50, 100, 200 kHz.
- Analog I/O: 2 channels in, 2 channels out
- Digital I/O: 16 channels configurable
RA16 Medusa Base Station
The RA16 is a single core Digital Sound Processor that can be interfaced with a multichannel ADC preamplifier like the RA8GA or RA4PA.
- Analog Input: up to 8 channels with external preamp via optical link
- Analog output: stereo via 6.35mm jack
- Digital I/O: input 1 bit (trigger), output 16 bits
- Sampling rate: 48.828 kHz maximum
RA8GA Adjustable Gain Preamplifier
The RA8GA is a 8 channel ADC device with adjustable preamp. It must be connected to the optical link input of a medusa base station (RA16) or RZ6. The RA8GA has female Dsub25 connector for the analog inputs. It is meant to be used with the PP16 patch panel with BNC connectors.
Specification | Value |
---|---|
A/D | 8-channels 16-bit PCM |
Maximum Voltage In | Variable gain settings allow ±10 V, ±1 V, or ±100 mV |
Frequency Response | DC - 7.5 kHz (2nd order 12 dB per octave) |
S/N (typical) | 70 dB (±1 V @ 1 kHz) at ±1 V Gain Setting |
THD (typical) | 0.01% |
A/D Sample Rate | 6, 12, or 25 kHz |
Cross Talk | <-70 dB (DC - Nyquist) |
Input Impedance | 10 kOhm |
Offset | <5 mV, ±10 V range <3 mV, ±1 V or ±100 mV range |
Clip warning light | signal > -6 dB. |
Build in low pass filter | Two-pole (12 dB per Octave) at 7.5 kHz. |
PA5 Progammable Attenuator
%todo
SA1 Stereo Amplifier
%todo
FO5 Fiber Optic Gigabit Zbus Interface
%todo
PO5e Fiber Optic Gigabit Computer Interface
%todo
PM2 Multiplexer
%todo
MS2 Monitor Speaker
%todo
ZBus for System 3
%todo
PS25F power supply
%todo