Difference between revisions of "Sphere lab technical info"
(Created page with "==Technical information== ===Sound booth=== <todo> ===Computer=== <todo> ===Electronics rack=== The electronics rack contains 10 TDT Zbus devices chassis, each containing two...") |
|||
Line 1: | Line 1: | ||
− | + | ==Sound booth== | |
− | |||
<todo> | <todo> | ||
− | + | ==Computer== | |
<todo> | <todo> | ||
− | + | ==Electronics rack== | |
The electronics rack contains 10 TDT Zbus devices chassis, each containing two TDT series 3 devices. Furthermore it contains the Field coil amplifier and the Femto LockinAmplifier as part of the head tracking electronics. | The electronics rack contains 10 TDT Zbus devices chassis, each containing two TDT series 3 devices. Furthermore it contains the Field coil amplifier and the Femto LockinAmplifier as part of the head tracking electronics. | ||
Line 85: | Line 84: | ||
[[File:image012.png|thumb|lower half of electronics rack]] | [[File:image012.png|thumb|lower half of electronics rack]] | ||
− | + | ==Sound system== | |
<todo> | <todo> | ||
====Minx12 speaker system==== | ====Minx12 speaker system==== | ||
Line 95: | Line 94: | ||
The PLC system (Panasonic FP2) consists of a PLC with several IO units and is can be programmed by via a special USB cable. | The PLC system (Panasonic FP2) consists of a PLC with several IO units and is can be programmed by via a special USB cable. | ||
− | + | ===Parts=== | |
*Trigger from Medusa base station | *Trigger from Medusa base station | ||
*PLC cabinet (houses PLC system) | *PLC cabinet (houses PLC system) | ||
Line 112: | Line 111: | ||
*Trigger converters (24V → 5V and 5V → 24V) | *Trigger converters (24V → 5V and 5V → 24V) | ||
− | + | ===Specifications=== | |
*FP2C2LJ FP2SH PLC | *FP2C2LJ FP2SH PLC | ||
**1 ms cycle time when <20.000 steps per cycle (120.000 steps max) | **1 ms cycle time when <20.000 steps per cycle (120.000 steps max) | ||
Line 137: | Line 136: | ||
**45 degree viewing angle | **45 degree viewing angle | ||
**Series resistor 1kOhm | **Series resistor 1kOhm | ||
− | + | **Actual current 10 mA @100% PWM |
Revision as of 09:58, 19 June 2023
Sound booth
<todo>
Computer
<todo>
Electronics rack
The electronics rack contains 10 TDT Zbus devices chassis, each containing two TDT series 3 devices. Furthermore it contains the Field coil amplifier and the Femto LockinAmplifier as part of the head tracking electronics.
The zBus Monitor program has control functions for the optical bus. It shows all the connected zBus chassis and the TDT devices that are installed in each chassis. It also shows the version number of the installed firmware on the devices.
The following table shows all the zBus chassis from top to bottom, with the installed devices.
Chassis | Front left | Front right | Back |
---|---|---|---|
Zbus | MS2 Monitor speaker | <empty> | |
Zbus | PA5 Attenuator | PA5 Attenuator | F05 optibit bus #1 |
Zbus | PA5 Attenuator | PA5 Attenuator | F05 optibit bus #2 |
Zbus | SA1 Speaker amplifier | SA1 Speaker amplifier | |
Zbus | RP2.1 Sound processor | RP2.1 Sound processor | F05 optibit bus #3 |
Zbus | PM2R Multiplexer | PM2R Multiplexer | |
Zbus | PM2R Multiplexer | PM2R Multiplexer | |
Zbus | PM2R Multiplexer | PM2R Multiplexer | |
Zbus | PM2R Multiplexer | PM2R Multiplexer | |
Zbus | RA16ba base station | RA8GA2 multi DAC | F05 optibit bus #4 |
The next list shows all other electronics in the lower half of the rack from top to bottom.
- PP16 Patch panel for Medusa Base Station (2x)
- Digital event recorder
- Junction box panel (connects to junction box in sound boot)
- Trigger distribution panel
- Femto Lockin amplifiers
- Tektronics Oscilloscope
- Field coil amplifier
- Rack main power switch
- Patch panel for XLR output to Tannoy speakers
Sound system
<todo>
Minx12 speaker system
<todo>
Tannoy speaker system
<todo>
LED system
The LED system consists of a cabinet inside the sound boot with a PLC system (programmable logic controller). The PLC can independently switch 256 channels. The channels connect to a two color LEDs (red and green) that are mounted on the speaker. The PLC cabinet has two test buttons that can light all LEDs at once. The PLC system (Panasonic FP2) consists of a PLC with several IO units and is can be programmed by via a special USB cable.
Parts
- Trigger from Medusa base station
- PLC cabinet (houses PLC system)
- PLC
- FP2SH CPU
- FP2 24V power supply
- FP2 12V power supply
- FP Web-Server
- FP2 input unit, 16 channels
- FP2 output unit, 64 channels (2x)
- FP2 multi I/O units (4 PWM channels)
- Break-out panel with 5 pole Binder connectors (128x)
- Binder cables (120x)
- LED mounting frames (120x)
- Red/Green LED’s (120x)
- Trigger converters (24V → 5V and 5V → 24V)
Specifications
- FP2C2LJ FP2SH PLC
- 1 ms cycle time when <20.000 steps per cycle (120.000 steps max)
- Measured cycle time 1.35 ms
- FP2Y64TJ output unit
- 24V NPN output
- Current 0.3A max
- Response time <0.3 ms
- FP2PXYPJ multi I/O unit
- PNP transistor
- PWM output 30 kHz
- Duty cycle 0-100% in steps of 1%
- Max output 800 mA
- FP2X16D2J input unit
- 16 inputs
- Current 8 mA @24V
- Response time <0.2 ms
- LED’s
- Bivar 5BC-3-CA-F (Common Anode)
- Red 625nm (FWHM = 25 nm)
- Green 568 nm (FWHM = 30 nm)
- Nominal current 20 mA
- Voltage drop 2.1V
- 45 degree viewing angle
- Series resistor 1kOhm
- Actual current 10 mA @100% PWM