Difference between revisions of "EM Field Head Tracking System specifications"
Jump to navigation
Jump to search
Line 49: | Line 49: | ||
{| class="wikitable" | {| class="wikitable" | ||
− | |||
! Time constant<br>Model „-H“ | ! Time constant<br>Model „-H“ | ||
! Hex switch<br>6dB/Octave | ! Hex switch<br>6dB/Octave | ||
! Hex switch<br>12dB/Octave | ! Hex switch<br>12dB/Octave | ||
|- | |- | ||
− | |||
| 300µs | | 300µs | ||
| 0HEX | | 0HEX | ||
| 8HEX | | 8HEX | ||
|- | |- | ||
− | |||
| 1ms | | 1ms | ||
| 1HEX | | 1HEX | ||
| 9HEX | | 9HEX | ||
|- | |- | ||
− | |||
| 3ms | | 3ms | ||
| 2HEX | | 2HEX | ||
| AHEX | | AHEX | ||
|- | |- | ||
− | |||
| 10ms | | 10ms | ||
| 3HEX | | 3HEX | ||
| BHEX | | BHEX | ||
|- | |- | ||
− | |||
| 30ms | | 30ms | ||
| 4HEX | | 4HEX | ||
| CHEX | | CHEX | ||
|- | |- | ||
− | |||
| 100ms | | 100ms | ||
| 5HEX | | 5HEX | ||
| DHEX | | DHEX | ||
|- | |- | ||
− | |||
| 300ms | | 300ms | ||
| 6HEX | | 6HEX | ||
| EHEX | | EHEX | ||
|- | |- | ||
− | |||
| 1s | | 1s | ||
| 7HEX | | 7HEX |
Revision as of 14:14, 29 November 2023
Introduction
The EM Field Head Tracking System is used for recording head movements of the subject during a trial. The system consists of the following parts:
- Three oscillators at roughly 45, 55 and 65 kHz or 50, 60, 70 kHz for X, Y and Z direction.
- Three signal amplifiers that amplify the oscillator signals.
- six field coils that make a cube of roughly 3x3x3 meters.
- A small pick-up coil mounted on a glasses frame that is worn by the subject during the experiment.
- Three lock-in amplifiers that split the signal from the pick-up coil in the 3 frequency components and measures their envelope. They output analog signals between -10V and +10V.
- An acquisition device converts the analog output signals from the lock-in amplifiers to digital.
- A RZ6 multi I/O processor that records the acquired data.
- Software that converts the recorded data to head orientations using calibration data.
The following table contains the channel convention with respect of the three directions:
Channel | Frequency | Orientation | Color | Polarity |
---|---|---|---|---|
Channel 1 | 45/50 kHz | Frontal | Red | Front = positive; Back = negative |
Channel 2 | 55/60 kHz | Vertical | Blue | Right = positive; Left = negative |
Channel 3 | 65/70 kHz | Horizontal | Yellow | Up = positive; Down = negative |
Parts
Lock-in amplifiers
Three Femto LIA-BV-150-H Lock-in amplifiers are used in order to separate the Frontal, Horizontal and Vertical signals from the pick-up coil. Each Lock-in amplifier has a Sinus Oscillator Module (SOM-1) that is used both as Field Modulation Signal and as Reference Signal. The Lock-in amplifier settings are set at the front of the device (see picture), except for the modulation/reference frequency which is set inside on the SOM-board.
The front panel of each LIA has four dipswitches:
- Mode: 1f means that the fundamental harmonic is measured (2f = second harmonic signal)
- Dyn.Res.: For dynamic reserve the settings Low (L) is preferred because the signal is normally large in comparison with the noise.
- PLL: For the Phase Lock Loop (PLL) the settings Slow (S) is preferred because the frequency is at a fix value. The phase is more stable in this setting.
- Ref.Thr.: Reference Threshold is set to 0V for analog signals (2V is for digital signals).
Time constant Model „-H“ |
Hex switch 6dB/Octave |
Hex switch 12dB/Octave |
---|---|---|
300µs | 0HEX | 8HEX |
1ms | 1HEX | 9HEX |
3ms | 2HEX | AHEX |
10ms | 3HEX | BHEX |
30ms | 4HEX | CHEX |
100ms | 5HEX | DHEX |
300ms | 6HEX | EHEX |
1s | 7HEX | FHEX |
Oscillators
Amplifiers
Resonance circuits
Wall coils
Pick-up coil
Lock-in amplifiers
ADC's
recording device
(RZ6 or RA16)