Three Femto LIA-BV-150-H Lock-in amplifiers are used in order to separate the Frontal, Horizontal and Vertical signals from the pick-up coil. Each Lock-in amplifier has a Sinus Oscillator Module (SOM-1) that is used both as Field Modulation Signal and as Reference Signal. The Lock-in amplifier settings are set at the front of the device (see picture), except for the modulation/reference frequency which is set inside on the SOM-board.
%todo
+
 
 
===Oscillators===
 
===Oscillators===
 
===Amplifiers===
 
===Amplifiers===

Revision as of 14:03, 29 November 2023

Schematic of the Sphere Head Tracking System

Introduction

The EM Field Head Tracking System is used for recording head movements of the subject during a trial. The system consists of the following parts:

  • Three oscillators at roughly 45, 55 and 65 kHz or 50, 60, 70 kHz for X, Y and Z direction.
  • Three signal amplifiers that amplify the oscillator signals.
  • six field coils that make a cube of roughly 3x3x3 meters.
  • A small pick-up coil mounted on a glasses frame that is worn by the subject during the experiment.
  • Three lock-in amplifiers that split the signal from the pick-up coil in the 3 frequency components and measures their envelope. They output analog signals between -10V and +10V.
  • An acquisition device converts the analog output signals from the lock-in amplifiers to digital.
  • A RZ6 multi I/O processor that records the acquired data.
  • Software that converts the recorded data to head orientations using calibration data.

The following table contains the channel convention with respect of the three directions:

Channel Frequency Orientation Color Polarity
Channel 1 45/50 kHz Frontal Red Front = positive; Back = negative
Channel 2 55/60 kHz Vertical Blue Right = positive; Left = negative
Channel 3 65/70 kHz Horizontal Yellow Up = positive; Down = negative

Parts

Lock-in amplifiers

Three Femto LIA-BV-150-H Lock-in amplifiers are used in order to separate the Frontal, Horizontal and Vertical signals from the pick-up coil. Each Lock-in amplifier has a Sinus Oscillator Module (SOM-1) that is used both as Field Modulation Signal and as Reference Signal. The Lock-in amplifier settings are set at the front of the device (see picture), except for the modulation/reference frequency which is set inside on the SOM-board.

Oscillators

Amplifiers

Resonance circuits

Wall coils

Pick-up coil

Lock-in amplifiers

ADC's

recording device

(RZ6 or RA16)