Difference between revisions of "EM Field Head Tracking System specifications"
Jump to navigation
Jump to search
(→Parts) |
|||
Line 1: | Line 1: | ||
+ | [[File:SphereHeadTrackingSystem.png|thumb|Schematic of the Sphere Head Tracking System]] | ||
==Introduction== | ==Introduction== | ||
− | The | + | The EM Field Head Tracking System is used for recording head movements of the subject during a trial. The system consists of the following parts: |
+ | *Three oscillators at roughly 45, 55 and 65 kHz or 50, 60, 70 kHz for X, Y and Z direction. | ||
+ | *Three signal amplifiers that amplify the oscillator signals. | ||
+ | *six field coils that make a cube of roughly 3x3x3 meters. | ||
+ | *A small pick-up coil mounted on a glasses frame that is worn by the subject during the experiment. | ||
+ | *Three lock-in amplifiers that split the signal from the pick-up coil in the 3 frequency components and measures their envelope. They output analog signals between -10V and +10V. | ||
+ | *An acquisition device converts the analog output signals from the lock-in amplifiers to digital. | ||
+ | *A RZ6 multi I/O processor that records the acquired data. | ||
+ | *Software that converts the recorded data to head orientations using calibration data. | ||
+ | The following table contains the channel convention with respect of the three directions: | ||
+ | |||
+ | {| class="wikitable" | ||
+ | ! Channel | ||
+ | ! Frequency | ||
+ | ! Orientation | ||
+ | ! Color | ||
+ | ! Polarity | ||
+ | |- | ||
+ | | Channel 1 | ||
+ | | 45 kHz | ||
+ | | Frontal | ||
+ | | Red | ||
+ | | Front = positive; Back = negative | ||
+ | |- | ||
+ | | Channel 2 | ||
+ | | 55 kHz | ||
+ | | Vertical | ||
+ | | Blue | ||
+ | | Right = positive; Left = negative | ||
+ | |- | ||
+ | | Channel 3 | ||
+ | | 65 kHz | ||
+ | | Horizontal | ||
+ | | Yellow | ||
+ | | Up = positive; Down = negative | ||
+ | |} | ||
==Parts== | ==Parts== | ||
− | [[ | + | ===Lock-in amplifiers=== |
+ | [[File:RobotLab_Lockin_Settings.JPEG|300px|thumb|right|Lock-in amplifier settings]] | ||
+ | %todo | ||
===Oscillators=== | ===Oscillators=== | ||
===Amplifiers=== | ===Amplifiers=== |
Revision as of 14:48, 23 November 2023
Introduction
The EM Field Head Tracking System is used for recording head movements of the subject during a trial. The system consists of the following parts:
- Three oscillators at roughly 45, 55 and 65 kHz or 50, 60, 70 kHz for X, Y and Z direction.
- Three signal amplifiers that amplify the oscillator signals.
- six field coils that make a cube of roughly 3x3x3 meters.
- A small pick-up coil mounted on a glasses frame that is worn by the subject during the experiment.
- Three lock-in amplifiers that split the signal from the pick-up coil in the 3 frequency components and measures their envelope. They output analog signals between -10V and +10V.
- An acquisition device converts the analog output signals from the lock-in amplifiers to digital.
- A RZ6 multi I/O processor that records the acquired data.
- Software that converts the recorded data to head orientations using calibration data.
The following table contains the channel convention with respect of the three directions:
Channel | Frequency | Orientation | Color | Polarity |
---|---|---|---|---|
Channel 1 | 45 kHz | Frontal | Red | Front = positive; Back = negative |
Channel 2 | 55 kHz | Vertical | Blue | Right = positive; Left = negative |
Channel 3 | 65 kHz | Horizontal | Yellow | Up = positive; Down = negative |
Parts
Lock-in amplifiers
%todo
Oscillators
Amplifiers
Resonance circuits
Wall coils
Pick-up coil
Lock-in amplifiers
ADC's
recording device
(RZ6 or RA16)